Specification of ADA 5000 Automatic Distillation Analyzer
Conforms to the specifications of: ASTM D86, D285, D850, D1078, D4737; D189 Section 10; DIN 51751; ISO 3405; IP 123; IP 195; JIS K2254l; NF M 07-002
Temperature Distillation Range: 0 to 450°C (±0.1°C accuracy)
Condenser: -5 to 60°C (±0.1°C accuracy); closed loop system
Receiver Chamber: 0 to 60°C (±0.1°C accuracy)
Distillation Parameters:
Distillation Rate: 2 to 15mL/min in 0.1mL increments, user selectable
Receiver Volume: 0 to 100mL (±0.01mL accuracy) by photoelectric infrared detection of meniscus by level following system utilizing a precision stepper motor and a special calibrated glass receiver; automatic calibration of evaporated loss volume and automatic volume calibration system ensures highest accuracy
Barometic Pressure: Automatic barometric correction utility with automatic sensor, range 550 to 900 mm Hg (±1 mm Hg accuracy)
Dry Point Detection: Automatic dry point detection board is included with standard equipment and only requires a dry point sensor, 200mL flask and PTFE plug for ASTM D850 and D1078 tests.
Environment: Operates at 0 to 25°C (113°F)
Untuk informasi terbaik hubungi:
Ibu WATY
Mobile: 081376851098